CVE intelligence and bounded remediation

CVE-2014-5434 — Baxter Sigma Spectrum Infusion System Firmware security vulnerability

Critical CVSS 9.8

Baxter SIGMA Spectrum Infusion System version 6.05 (model 35700BAX) with wireless battery module (WBM) version 16 has a default account with hard-coded credentials used with the FTP protocol. Baxter asserts no files can be transferred to or from the WBM using this account. Baxter has released a new version of the SIGMA Spectrum Infusion System, Version 8, which incorporates hardware and software changes.

Severity
Critical
CVSS
9.8 (3.0)
Published
2019-03-26
CISA KEV
Not currently listed
Ecosystem
operating-system
Weaknesses
CWE-259, CWE-798

Affected products

  • baxter / sigma_spectrum_infusion_system_firmware / 6.05

Matched remediation archetype

Authentication bypass and missing authentication

This catalog composition supplies bounded fallback guidance. Explicitly reviewed curated workflows load with the complete record below.

Check exposure

  • Map every affected endpoint and protocol path, including alternate ports, legacy routes, recovery flows, service accounts, and machine-to-machine access.
  • Confirm which deployments enable the affected authentication mode and whether the interface is reachable from untrusted networks.
  • Review session, token, credential, and proxy trust configuration without attempting account takeover.

Remediate safely

  • Apply the supported fix and centralize fail-closed authentication before protected request handling.
  • Remove default or embedded credentials, rotate affected secrets and sessions, and bind authentication decisions to the intended audience and channel.
  • Add negative tests for alternate routes, malformed or absent credentials, recovery flows, and proxy-derived identity.

Authoritative sources

Complete CVE record and remediation plan

The detailed catalog view below loads this exact record, its source evidence, and the full seven-phase agentic change plan.